Absorbance and optical density (OD)
Why “OD” is an older name for absorbance, and why OD600 readings of cultures are mostly scattering.
An older name for the same logarithm
Optical density, or OD, is an older term for absorbance. IUPAC lists it as an obsolete synonym and discourages its use, but it remains common on instruments, in protocols and in biology.
Whatever the label, the arithmetic of a decadic reading is the same: a reading of 1 means 10% of the light reached the detector, and a reading of 2 means 1%.
When light is scattered, not absorbed
A spectrophotometer measures how much light fails to reach its detector. In a clear solution that loss is mainly absorption. In a suspension, such as a bacterial culture read at 600 nm (“OD600”), much of the loss comes from light scattered away from the detector.
IUPAC calls a quantity that also includes scattering and luminescence attenuance, rather than absorbance. How much scattered light misses the detector depends on the instrument’s optics, so turbidity readings from different instruments may not agree, and a Beer–Lambert coefficient does not describe them. Relating OD600 to cell numbers needs an empirical calibration for the instrument and organism.
What the tools on this site assume
The converter’s arithmetic between %T and A applies to any decadic reading. The Beer–Lambert calculator and the visual explainer assume absorption by a homogeneous, non-scattering sample, so they do not describe the turbidity of a suspension.