Absorbance glossary
Short, plain-English definitions of the terms used across the tools and guides.
- Absorbance (A)
- A dimensionless measure of how much a sample reduces the radiant power of a beam: A = log₁₀(P₀ / P) = −log₁₀(T). This site uses decadic (base-10) absorbance. An absorbance of 1 means 10% of the light is transmitted. What does absorbance measure? ↗
- Absorption
- The physical process in which matter takes up radiation. Absorption is the process; absorbance is a defined number used to quantify its effect on a beam.
- Amount concentration (molarity)
- Amount of substance per volume of solution, usually in mol/L, written M. Older texts call it molarity. 1 mM = 10⁻³ mol/L and 1 µM = 10⁻⁶ mol/L. Path length and concentration units ↗
- Attenuance
- A quantity like absorbance that also includes losses from scattering and luminescence; formerly called extinction. Turbid samples, such as cell suspensions, are better described by attenuance than by absorbance. Absorbance and optical density ↗
- Beer–Lambert law
- For an idealised homogeneous, non-scattering sample, absorbance is proportional to path length and concentration: A = ε × l × c. Real samples can depart from it. Using the Beer–Lambert law ↗
- Blank
- A sample containing everything except the substance being measured, such as the solvent and reagents, used as the zero reference. Measured absorbance is relative to the blank, which is why slightly negative readings can occur.
- Calibration curve (standard curve)
- A graph of an instrument response, such as absorbance, against the known concentrations of standards. The fitted line is inverted to estimate unknown concentrations. Calibration curve builder ↗
- Cuvette (cell)
- The transparent container that holds a sample in the light beam. Its internal width sets the optical path length; 1 cm is a common size.
- Decadic and Napierian absorbance
- Decadic absorbance uses base-10 logarithms. Napierian absorbance uses natural logarithms and is ln 10 ≈ 2.303 times larger for the same sample. Spectrophotometers usually report, and this site uses, decadic absorbance.
- Dilution factor (D)
- The original concentration divided by the diluted concentration: D = C₁ / C₂ = V₂ / V₁. Diluting 1 mL to a total of 10 mL gives D = 10. Dilution calculator ↗
- Extrapolation
- Estimating a value outside the range covered by the data, such as a concentration above the highest standard. The calibration has not been checked there, so the calibration builder flags these estimates.
- Intercept (b)
- The fitted response at zero concentration. In a calibration it reflects any blank offset. The calibration builder fits it rather than forcing the line through zero. Read a linear calibration curve ↗
- Mass concentration
- Mass of a substance per volume of solution, such as mg/mL or g/L (1 mg/mL = 1 g/L). Dividing by the molar mass converts it to amount concentration. Molarity calculator ↗
- Molar absorptivity (ε)
- The constant in A = ε × l × c for a given substance, wavelength and set of conditions, usually in L·mol⁻¹·cm⁻¹. IUPAC prefers the name molar absorption coefficient; extinction coefficient is an older name. Molar absorptivity explained ↗
- Molar mass (M)
- Mass per amount of substance, in g/mol. It links mass and amount: n = m / M. Use the molar mass of the exact form weighed, including any water of hydration. Molarity calculator ↗
- Optical density (OD)
- An older name for absorbance whose use IUPAC discourages. In microbiology, OD600 readings of cultures largely reflect light scattering rather than absorption. Absorbance and optical density ↗
- Ordinary least squares (OLS)
- A fitting method that chooses the line minimising the sum of squared vertical differences between the data and the line. It treats the x values, here the standard concentrations, as exact.
- Path length (l)
- The distance light travels through the sample, usually in centimetres. In the Beer–Lambert model, doubling the path length doubles the absorbance.
- R² (coefficient of determination)
- For a least-squares line with an intercept, the fraction of the variation in the response that the line explains, from 0 to 1. A high R² does not show that a straight line is the right model or that a method is valid.
- Residual
- A measured value minus the value predicted by the fitted line. A pattern in the residuals, such as a curve, can reveal a problem that R² hides.
- Serial dilution
- A sequence of dilutions in which each solution is made from the previous one, usually by the same factor. The overall dilution is the product of the step factors, and errors carry forward. Preparing calibration standards ↗
- Standard (calibration standard)
- A solution of known concentration, measured to build a calibration curve.
- Stray light
- Light outside the intended wavelength band, or light that bypasses the sample, that still reaches the detector. It limits the highest absorbance an instrument can measure accurately.
- Transmittance (T, %T)
- The fraction of incident radiant power that passes through the sample, T = P / P₀. Percentage transmittance is 100 × T, and T = 10⁻ᴬ. Absorbance to transmittance converter ↗
- Wavelength (λ)
- The position in the spectrum of the light used for a measurement, usually given in nanometres. Absorbance and ε depend on wavelength, so a measurement is made at a stated wavelength.
Original definitions written for this site. Sources are listed with each guide; see the IUPAC Gold Book for formal terminology.