Preparing calibration standards by dilution
Plan a set of standards from one stock, and choose between serial and parallel dilution.
Start from C₁V₁ = C₂V₂
Diluting adds solvent but not solute, so the amount taken from the stock equals the amount in the final solution. To make 10 mL of a 10 µM standard from a 100 µM stock: V₁ = 10 µM × 10 mL / 100 µM = 1 mL. Measure 1 mL of stock and make up to 10 mL in total.
“Make up to 10 mL” and “add 10 mL” are different preparations. Volumes are not always exactly additive, so for accurate work make each solution up to the mark in a volumetric flask.
Parallel or serial?
In a parallel dilution, each standard is made directly from the stock. An error in one standard does not carry into the next, but the lowest standards may need awkwardly small stock volumes.
In a serial dilution, each solution is made from the previous one, often with the same factor at each step. This spans wide ranges with convenient volumes, but an error in any transfer is carried into every later solution. The overall dilution is the product of the step factors.
Plan the set before you pipette
Choose concentrations that span the range you need, bracket the samples you expect to measure, and include a blank. Equal spacing across the range is a practical default. The NIST/SEMATECH e-Handbook recommends at least five reference standards for a linear calibration curve, with repeated measurements of each.
Where practical, prepare standards in the same solvent or matrix as your samples, so that the blank and the response are comparable. These are general planning points, not a validated preparation protocol for any particular method.