Dilution calculator
Solve C₁V₁ = C₂V₂ for any one value, plan a serial dilution, or work out how to make a set of calibration standards from a single stock.
Enter your values or load an example.
The amount of solute stays the same
Diluting adds solvent, not solute, so the amount in the stock you take equals the amount in the final solution: C₁ × V₁ = C₂ × V₂. The dilution factor is D = C₁ / C₂ = V₂ / V₁.
Units are converted before calculating, so you can take a stock in M and a target in mM or µM. Mass concentrations (such as mg/mL) work the same way, but cannot be mixed with molar units without a molar mass.
Dilution factors without the mix-ups ↗Arithmetic, not a protocol
These plans assume the solute is conserved and fully mixed, and that volumes are measured as entered. Make each solution up to its final volume rather than relying on volumes being additive. Pipette accuracy, adsorption and stability are not modelled.
Planning standards for a calibration curve? Spread them across the range you need to measure, bracket your expected samples and include a blank.
Preparing calibration standards ↗Common questions
How do I use C₁V₁ = C₂V₂?
C₁ and V₁ are the concentration and volume of the stock you take; C₂ and V₂ are the concentration and total volume of the diluted solution. Because the amount of solute is unchanged by dilution, C₁ × V₁ = C₂ × V₂. Enter any three values to find the fourth. To make 100 mL of 0.1 M from a 1 M stock: V₁ = 0.1 × 100 / 1 = 10 mL.
How much diluent should I add?
Make the solution up to the final volume V₂. The difference V₂ − V₁ is a good guide to the diluent needed, but volumes are not always exactly additive, so for accurate work add the stock to a volumetric flask and fill to the mark.
What is a dilution factor?
The dilution factor is the stock concentration divided by the diluted concentration, C₁ / C₂, which equals V₂ / V₁. Diluting 1 mL of stock to a total of 10 mL is a ten-fold dilution (D = 10). In a serial dilution, the overall factor is the product of the factors at each step.
Serial or parallel dilution: which is better for standards?
Serial dilution (each solution made from the previous one) is convenient for wide ranges, such as ten-fold steps, but any volume error is carried into every later tube. Parallel dilution (each standard made directly from the stock) keeps errors independent, at the cost of measuring some small volumes.
Can I mix mg/mL and molar units?
Not in one dilution: C₁ and C₂ must both be amount concentrations (M, mM, µM, nM) or both be mass concentrations (g/L, mg/mL, mg/L, µg/mL, µg/L). Converting between the two needs the molar mass of the substance, which the molarity calculator handles.