From your standards to a clear picture. Fit absorbance against concentration, inspect the line and estimate an unknown. All data stays in this browser.

01 · Your standards

Synthetic teaching data · four illustrative standards.

#Concentration (µM)AbsorbanceRemove
Import a CSV or paste a table

Up to 2 MB and 5,000 standards. The first nonblank row must contain column headers. Use decimal points, not decimal commas.

Download the synthetic sample CSV ↓

02 · Your calibration

Synthetic teaching data

Scroll the graph horizontally, or open the values table below.

Slope (A / µM)
Intercept (A)
R²
Standards
Inspect fitted values & residuals

Residual = measured absorbance − fitted absorbance. Replicates are retained.

Original standards and results
ConcentrationAbsorbanceFittedResidual

03 · Estimate an unknown

Calculation summary from www.absorbance.net. This is an educational and general calculation, not a certificate or laboratory method validation. Full-precision numerical results are available in the CSV export.

Know what the line is doing

This tool fits absorbance (y) against known concentration (x) using ordinary least squares with an intercept: A = m × c + b. It never forces the intercept to zero, averages replicates, removes outliers or applies a blank correction.

Unknown concentrations are calculated by inverting that same fitted line: c = (A − b) / m. Any dilution factor applies to the measured-solution estimate, not the absorbance.

How to read the graph ↗Plan your standards with the dilution calculator ↗

Look beyond R²

Standards are treated as fixed/exact, the linear model must be appropriate, and measurement variability is assumed reasonably comparable across standards. Concentration uncertainty is not modelled. High R² does not prove method validity.

Negative measured absorbance is retained. Negative estimates and extrapolation are flagged. Uncertainty is not calculated in this first release. Source: NIST model validation.

Preparing calibration standards ↗

Common questions

How many standards do I need for a calibration curve?

This tool needs at least three rows with two distinct concentrations to fit a line, but that is a software minimum, not a recommendation. The NIST/SEMATECH e-Handbook of Statistical Methods recommends at least five reference standards for a linear calibration curve, spread across the range of interest, with repeated measurements of each.

Is a high R² enough to trust a calibration?

No. R² summarises how much of the variation in absorbance the straight line explains, but curvature, an influential point or uneven scatter can exist even when R² is above 0.99. Inspect the graph and the residuals as well.

Should I force the calibration line through zero?

This builder always fits an intercept. Forcing the line through the origin assumes the blank response is exactly zero; if it is not, the slope and every estimate are biased. A fitted intercept shows the offset instead of hiding it.

How is an unknown concentration calculated from the curve?

The fitted line A = m × c + b is inverted: c = (A − b) / m. If the sample was diluted before measurement, multiply by the dilution factor to estimate the original concentration. Estimates outside the range of your standards are flagged as extrapolation.

Can I use the graph in a report or assignment?

Yes. Export the graph as SVG (scalable, for documents and slides) or PNG, the full-precision results as CSV, or print a summary to PDF. Exports state the model, units and assumptions. They are educational calculations, not validated laboratory reports.